Compliance - Model CDT-240 V2 8A - Capacitor Discharge Tester
From Capacitive Discharge Testers
For IEC Testing of residual voltage after plug disconnect. The CDT-240 V2 disconnects the line at the voltage peak and analyzes results to the requirements of various Standards, allowing quick results. Scope output for viewing decay waveform. Presets for EN 61010, EN 60065, EN 60950, EN 60601-1 and EN60335. Recommendations for Pluggable Type B equipment. NIST calibration supplied; A2LA calibration available. 15A option for testing large equipment.
-
Most popular related searches
Product Details
For IEC Testing of residual voltage after plug disconnect. The CDT-240 V2 disconnects the line at the voltage peak and analyzes results to the requirements of various Standards, allowing quick results. Scope output for viewing decay waveform. Presets for EN 61010, EN 60065, EN 60950, EN 60601-1 and EN60335. Recommendations for Pluggable Type B equipment. NIST calibration supplied; A2LA calibration available. 15A option for testing large equipment.
Features
- Accurate; in accordance with CTL voltage and time specifications.
- Disconnects input test voltage at positive or negative peak, normal and reversed polarity
- Tests Line-Ground, Line-Neutral and Neutral-Ground.
- Micro-controller Technology, one button test, easy setup, LCD display indicates test results.
- Maximum Rated Test Current 8A, 15A optional.
- Ergonomically designed for safety, speed and efficiency.
- Designed for product development, production and laboratory use.
- Rugged construction that can withstand a 6' drop with no damage.
- Cables, manual and calibration certificate included.
- One year warranty.
Technical Specification
- Min/Max Test Voltage: 90/264 Vrms
- Max Test Current: 8A, 15A option available
- Fuse: 10A, 250V, time delay, 1-1/4' x 1/4'
- Probe Impedance: > 100 Mohm, < 25pF
- AC Peak Detection Accuracy: EUT disconnect within 5% of supply voltage peak
- Test Time Range: 0.5 - 10.0 Seconds
- Voltage Setting Range: +/-30 to 388V
- Voltage Setting Resolution: 1V
- Test Time Accuracy: +/-5%
- Voltage Meter Accuracy: +/-2% of reading +/-0.8V
- Frequency: 50-60Hz
ELECTRICAL:
- Rated Input Voltage: 110-240 V
- Frequency: 50/60 Hz
- Fuse: 2A, 250V, time delay, 5x20 mm
- ENVIRONMENTAL:
- Operation Temperature: 15 - 40 C
- Relative Humidity Range: 0 - 90 % Non-Condensing
SAFETY:
- (XI) Safety Interlock: 2 pin terminal block on rear panel, shorted for tester operation
- GENERAL:
- Dimms: 9-1/4'W x 4-3/4'H x 10'D
- Wt: 5.8lbs
- BNC Output: Referenced to ground, 100:1 of measured voltage
Customer reviews
No reviews were found for Compliance - Model CDT-240 V2 8A - Capacitor Discharge Tester. Be the first to review!