Model MC6 - Multi-Cal Benchtop Calibrator
From Test Equipment
Model MC6 Multi-Cal Benchtop Calibrator performs a wide variety of simple and complex pressure based measurement, test, and calibration operations. Modular sensor design allows user to select pressure measurement range for application flexibility. Calibrator can accommodate up to two interchangeable pressure modules (sold separately) in any combination of range or accuracy. Simultaneously display two separate measurements on the two line, alphanumeric display. Readings can be displayed in a choice of twelve preprogrammed engineering units or any single user-defined unit. Calibrator features min/max recall, hi/lo alarm, percentage of full scale pressure readings, mA/voltage measurement, leak rate and pressure decay measurement, switch testing capabilities, including trip point and dead band, and velocity/volume flow rates. Quickly document calibration procedures using the data logging feature which stores up to 384 sets of pressure and time/date stamped measurements.
-
Most popular related searches
Specifications
Service: Clean, dry, nonconductive, noncorrosive gases.
Accuracy: Differential Pressure Modules: ±0.06% FS, Gauge Pressure Modules: ±0.05% FS, Voltage Input: ±0.025% FS @ 1/10 VDC, ±0.10% FS @ 0/30 VDC, Current Input: ±0.03% FS @ 1/20 mA, ±0.05% FS @ 0/50 mA.
Sensitivity: ±0.002% of span with dampening 1 part in 50,000 (max.).
Repeatability: Ranges ≤0/2 psi: ±0.05% of span, Ranges ≥0/5 psid: ±0.02% of span.
Output: RS232 serial interface, 9-pin.
Alarm Output: SPST form C 110 VDC, 120 VDC (max.), 1 A (mas.), 30 W, 62.5 VA (resistive).
Display: Alphanumeric LCD, 0.37' (9.5 mm) height per line, 2 lines, 16 characters/line.
Display Update: 100 msec.
Ambient Operation Temperature: 32 to 120°F (0 to 49°C).
Storage Temperature: -4 to 158°F (-20 to 70°C).
Electrical Connections: Starndard banana jacks.
Power Supply: Internal: two 9 VDC alkaline batteries,External: AC adapter 9 VDC, 500 mA.
Battery Life: 30 hours (approximate).
Engineering Units: in H2O, psi, in Hg, kPa, mbar, bar, cm H2O, mm Hg, ft SW, MPa, mm H2O, kg cm² and user-defined.
Overpressure: Differential Pressure Modules: 50 psi positive direction, 15 psi negative direction, Gauge Pressure Modules: 2x range (0/5 psi to 0/1000 psi).
Temperature Compensation: 20 to 120°F (-7 to 49°C).
Temperature Error: Maximum of ±0.004% of span per °F over compensated range for zero and span.
Temperature Effect Electrical Measurement: ±0.001% of span per °F over compensated range.
Dampening: (Measurement averaging) programmable from 0 to 16 consecutive readings.
Baud Rate: 300, 1200, 2400, or 9600, selectable.
Computer Requirements: IBM compatible 386 or above and Windows® 3.1 or later with 2mB RAM and 2mB hard drive disk space, one serial port.
Housing Material: ABS plastic.
Weight: Calibrator: 3 lb (1.4 kg), Pressure Module: 0.5 lb (0.3 kg).
Agency Approvals: CE.
Customer reviews
No reviews were found for Model MC6 - Multi-Cal Benchtop Calibrator. Be the first to review!