Triad Magnetics - Medical Grade Isolation Power Transformer MD-250-E
From Power Transformers - Medical Grade Isolation Power Transformers
Electrical Specifications at 25 Degree Celsius (ºC) Temperature
Maximum Power: 250 V·A
Input Voltage1: 240 V
Output Voltage2: 120 V
Input Voltage Tolerance (Power Cord): ±10 %
Frequency (Power Cord): 50 Hz, 60 Hz
Note for Output Voltage: Alternating Current (AC) Output Voltage (Receptacle) at Maximum 2.08 A.
Typical Voltage Regulation from Full Load to No Load: 8 %
Alternating Current (AC) Input to Output Voltage (Hi-Pot): 3000 V
Alternating Current (AC) Input and Output to Enclosure Voltage (Ground) (Hi-Pot): 1500 V
Typical Efficiency at Full Load: 92 %
Operating Ambient Temperature: Not to Exceed 60 ºC
Leakage Current: Less Than 50 µA
Typical Leakage Current: 10 µA
Fusing: Replace only with 1.8 A, 250 VAC, Slow Blow, UL Listed Fuses.
Specifications
Connections
Input Connections: 11 ft. Medical Grade Power Cord with NEMA 6-15P Plug.
Output Connections: Duplex Receptacle. NEMA 5-15R.
Note
Note: Medical isolation transformer box is certified to UL 60601-1, 2nd edition medical standard and CSA-C22.2.No. 601.1M90, 2005
UL: (E352717), UL60601-1, 2nd edition and CAN/CSA-C22.2 No. 601.1M90, 2005
Not for use in life critical application.
Customer reviews
No reviews were found for Triad Magnetics - Medical Grade Isolation Power Transformer MD-250-E. Be the first to review!