Triad Magnetics

Triad Magnetics

- Medical Grade Isolation Power Transformer MD-250-E

From

0
The MD-250-E is a medical grade isolation power transformer designed with extensive safety and reliability features suitable for medical applications. Operating at a frequency of 50 to 60 Hz, it has an impressive capacity to handle up to 250 VA of power. It features an input voltage range of 120 to 240 V with a tolerance of ±10%, ensuring versatile usage in various settings. The transformer includes robust electrical isolation with a high-potential (Hi-Pot) input to output voltage rating of 3000 V and a ground rating of 1500 V. With an efficiency of 92% at full load, it operates in ambient temperatures not exceeding 60ºC. The leakage current is maintained below 50 µA to comply with medical safety standards. Constructed with a steel powder-coated enclosure, it also provides certified compliance to UL 60601-1 and CSA-C22.2 medical standards. The design includes rubber feet for horizontal mounting and comes equipped with a 11 ft medical grade power cord and duplex receptacle for user convenience.

Details

These medical grade isolation boxes are built using a toroidal transformer construction that inherently helps reduce stray fields, increases efficiency and minimizes size compared to traditional EI transformers. Built with Class F (155º) insulation system. The enclosure is made of steel and powder coated white.
 
 

Electrical Specifications at 25 Degree Celsius (ºC) Temperature

Maximum Power: 250 V·A

Input Voltage1: 240 V

Output Voltage2: 120 V

Input Voltage Tolerance (Power Cord): ±10 %

Frequency (Power Cord): 50 Hz, 60 Hz

Note for Output Voltage: Alternating Current (AC) Output Voltage (Receptacle) at Maximum 2.08 A.

Typical Voltage Regulation from Full Load to No Load: 8 %

Alternating Current (AC) Input to Output Voltage (Hi-Pot): 3000 V

Alternating Current (AC) Input and Output to Enclosure Voltage (Ground) (Hi-Pot): 1500 V

Typical Efficiency at Full Load: 92 %

Operating Ambient Temperature: Not to Exceed 60 ºC

Leakage Current: Less Than 50 µA

Typical Leakage Current: 10 µA

Fusing: Replace only with 1.8 A, 250 VAC, Slow Blow, UL Listed Fuses.

Specifications

Connections

Input Connections: 11 ft. Medical Grade Power Cord with NEMA 6-15P Plug.

Output Connections: Duplex Receptacle. NEMA 5-15R.

Note

Note: Medical isolation transformer box is certified to UL 60601-1, 2nd edition medical standard and CSA-C22.2.No. 601.1M90, 2005

UL: (E352717), UL60601-1, 2nd edition and CAN/CSA-C22.2 No. 601.1M90, 2005

 

Not for use in life critical application.

Customer reviews

No reviews were found for Triad Magnetics - Medical Grade Isolation Power Transformer MD-250-E. Be the first to review!